From the remote interface:
To configure the List voltage of output 1 with 1 V, 2 V, 3 V, 4 V and 5 V:
LIST:VOLT 1,2,3,4,5, (@1)
To configure the List current of output 1 with 0.1 A, 0.2 V, 0.3 A, 0.4 A and 0.5 A:
LIST:CURR 0.1,0.2,0.3,0.4,0.5, (@1)
To configure all the List time of output 1 with 1 s:
LIST:DWELL 1,1,1,1,1, (@1)
To enable all the List BOST of output 1:
LIST:TOUT:BOST 1,1,1,1,1, (@1)
To disable all the List EOST of output 1:
LIST:TOUT:EOST 0,0,0,0,0, (@1)
To set the voltage mode of output 1 to List:
VOLT:MODE LIST, (@1)
To set the current mode of output 1 to List:
CURR:MODE LIST, (@1)
To set the trigger source of output 1 to Key/Immediate:
TRIG:SOUR IMM, (@1)
To enable output 1:
OUTP ON, (@1)
To initiate and run the list:
INIT (@1)