When exporting the data, specify the output you want to export the logged data.
Press
Save Path
to browse and specify the location where the export file will be placed. Use the front-panel
navigation keys to navigate through the list. The left and right arrows contract or expand a folder to hide or show its
files. Press
Select
once you have specified the path or
Cancel
to abort the changes made.
Press
File Name
to configure the filename. Use the keyboard to enter your desired filename in the File Name field.
Press
Back
to save or
Cancel
to abort the changes made.
Press
Export File
to export the file.
From the remote interface:
To program output sequence of five steps on output 1:
LIST:VOLT 1,2,3,4,5, (@1)
LIST:CURR 0.1,0.1,0.1,0.1,0.1, (@1)
LIST:DWEL 1,1,1,1,1, (@1)
LIST:COUNT 5, (@1)
LIST:STEP AUTO, (@1)
VOLT:MODE LIST, (@1)
CURR:MODE LIST, (@1)
To initiate the trigger system:
TRIG:SOUR BUS
INIT (@1)
To set up the data log for output 1:
SENS:DLOG:FUNC:VOLT 1, (@1)
SENS:DLOG:TIME 30
SENS:DLOG:PER 0.2
To initiate that data logger and specify the filename in which to save the data:
Keysight E36200 Series User's Guide
97
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...