Rear panel
If voltage drop in the load leads is a concern, connect the sense leads of output 1 and output 2 for 4-wire sensing as
shown in the figure on the right.
4-wire sense connection (E36233A and E36234A only)
SHOCK HAZARD
Turn off AC power before making rear panel connections. All wires and straps must be properly connected
with the terminal block screws securely tightened
The power supply includes built-in relays that connect or disconnect the ± sense terminals from their corresponding
± output terminals. As shipped, the sense terminals are internally connected to the output terminals. This is referred
to as 2-wire, or local sensing.
4-wire, or remote sensing improves the voltage regulation at the load by monitoring the voltage at the load rather
than at the output terminals. This automatically compensates for the voltage drop in the load leads, which is
especially useful for CV operation with load impedance that vary or have significant lead resistance. Because remote
sensing is independent of other power supply functions, it can be used regardless of how the instrument is
programmed. Remote sensing has no effect during CC operation.
38
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...