Specifying 2-Wire or 4-Wire Sense
When 4w mode is selected, ensure the load is connected to the rear panel output for accurate reading.
DO NOT USE both front and rear panel output concurrently. This may lead to inaccurate output reading.
You have the option to use the remote sense terminals on the rear panel to measure the voltage at the device under
test (DUT).
1.
Press
Source Settings
.
2.
Press
Sense 2w
or
Sense 4w
to select 2-wire sense or 4-wire sense.
3.
Press
Back
to exit and return to the meter-view display.
The 2w mode closes a relay within the power supply to short the output and sense connectors. This means that only
two wires are used, and 4-wire sense is disabled. The 4w mode opens the relay in order to separate the output and
4-wire sense inputs. By default, 2w is selected.
From the remote interface:
To set the remote sense relay to 4-wire sense at output 1:
VOLT:SENS EXT (@1)