Using the Tracking Operation (E36233A/E36234A Only)
The E36233 and E36234 models provide tracking outputs. In the track mode, the voltages from Output 1 and
Output 2 track each other for convenience in varying the symmetrical voltages needed by operational amplifiers and
other circuits using balanced positive and negative inputs. The state of the track mode is stored in volatile memory.
The track mode is always in off state even when power is being disconnected or after a remote interface reset.
1.
Set Output 1 to your desired voltage.
2.
Press
Output Settings
>
Operation Mode
>
Mode Tracking
. When track mode is enabled, Output 2 will be set
to the same voltage level as the Output 1 and vice versa. The current limit is independently set for each of the
outputs and is not affected by the track mode.
3.
Verify that both Output 1 and 2 track each other properly. You can verify from the front-panel display by com-
paring the voltage values of Output 1 and Output 2.
In the track mode, if the CC annunciator is lit, choose a higher current limit for the voltage supply.
From the remote interface:
To enable track mode:
OUTP:TRAC ON
74
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...