Using the Protection Function
Protection function
Each output has independent protection functions. A front panel status indicator will turn on when a protection
function is set. Protection functions are latching, which means that they must be cleared once they are set.
Of the following protection functions, only OV and OC are user-programmable.
–
OV
: Over-voltage protection is a hardware OVP whose trip level is a programmable value. The OVP is enabled by
default.
–
OC
: Over-current protection is a programmable function that can be enabled or disabled. When enabled, the
output will be disabled when the output current reaches the current limit setting.
–
OT
: Over-temperature protection monitors the temperature of each output and shuts down the output if any
temperature exceeds the maximum factory-defined limits.
Configuring protection
Protection functions are configured on the Source Settings window.
Press
Source Settings
to access the Source Settings window. Next, press
Protection
to access the protection
functions.
Keysight E36200 Series User's Guide
69