Rear outputs connections
SHOCK HAZARD
Turn off AC power before connecting wires to the rear panel. All wires and straps must be properly connected
with the terminal block screws securely tightened.
Disconnect the connector plug to make your wire connections. The connector plug accepts wires sizes from AWG
12 to AWG 30. The connector plug is 20 A rated for the E36231A, E36232A, and E36234A; and 41 A rated for the
E36233A. Wire sizes smaller than AWG 20 are not recommended. Connect the load wires to the + and - terminals.
Connect the sense wires to the +s and -s terminals.
Securely fasten all wires by tightening the screw terminals. Insert the connector plug into the back of the unit. A
chassis ground binding post is located next to the fan ventilation hole for ground connections.
Part number information for the connector kit is provided under
Wire sizing
FIRE HAZARD
Select a wire size large enough to carry short-circuit current without overheating (refer to the following table).
To satisfy safety requirements, load wires must be heavy enough not to overheat while carrying the short-cir-
cuit output current of the unit.
AWG
Suggested maximum
current (A)
1
mΩ/ft
mΩ/m
12
41
1.59
5.2
14
32
2.53
8.3
16
22
4.02
13.2
18
16
6.39
20.9
20
11
10.2
33.3
22
7
16.1
52.9
24
3.5
25.7
84.2
Notes:
1. Single conductor in free air at 30 °C with insulation
32
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...