Keysight M8070A Programming Guide
429
SCPI Command Reference
5
Description
This command sets the value till the sweep needs to be performed for that
particular parameter.
This query returns the current sweep value till the sweep is performed.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:ACQ:STO 'ParameterSweep1', 1
:PLUGin:PSWEep:ACQuisition:STEP[?]
Syntax
:PLUGin:PSWEep:ACQuisition:STEP 'Identifier', <NRf>
:PLUGin:PSWEep:ACQuisition:STEP? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
Description
This command sets the value as for how many steps user wants to sweep
the parameter between the current “Sweep From” and “Sweep To” range.
This query returns the current number of steps set for the acquisition.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:ACQ:STEP 'ParameterSweep1', 100
:PLUGin:PSWEep:ACQuisition:COMPared[?]
Syntax
:PLUGin:PSWEep:ACQuisition:COMPared 'Identifier', <NRf>
:PLUGin:PSWEep:ACQuisition:COMPared? 'Identifier',
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
<location-string>: ‘M*.DataIn*’
Description
This command allows the user to specify the number of compared bits;
that would be used as criteria for moving to the next measurement point in
case when the sample delay optimization is “NONE”.
This query returns current value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:ACQ:COMP 'ParameterSweep1', 1E+7