Keysight M8070A Programming Guide
127
SCPI Command Reference
5
:STATus:INSTrument:SALoss?
Syntax
:STATus:INSTrument:SALoss? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2' or ‘M1.System’
Return Range
0|1
Description
This query indicates if the symbol alignment has been lost. A true (1)
indicates a symbol alignment loss.
Example
:STAT:INST:SAL? ‘M1.DataIn1’
:STATus:INSTrument:ERRor?
Syntax
:STATus:INSTrument:ERRor? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2' or ‘M1.System’
Return Range
0|1
Description
This query indicates if an error condition has occurred. A true (1) indicates
an error condition.
Example
:STAT:INST:ERR? ‘M1.DataIn1’
:STATus:INSTrument:OVERload?
Syntax
:STATus:INSTrument:OVERload? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2' or 'M*.DataOut1' or 'M*.DataOut2'
or ‘M1.System’
Return Range
0|1
Description
This query indicates an overload condition exists at an input or output. A
true (1) indicates an overload condition.
Example
:STAT:INST:OVER? ‘M1.DataIn1’