202
Keysight M8070A Programming Guide
5
SCPI Command Reference
[:SOURce]:CONFigure Subnode
This subnode has the following SCPI structure:
This subnode has the following command:
[:SOURce]:CONFigure:MINTegration[?]
Syntax
[:SOURce]:CONFigure:MINTegration
‘identifier’,<NONe|MUX|DMUX|BOTH>[<data-location>[<visa-resource>]]
[:SOURce]:CONFigure:MINTegration? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.MuxMode’
<NONe|MUX|DMUX|BOTH>: Select mux/demux module.
<data-location>: (Optional) Specify location of data module. If omitted or
an empty string is specified, the first suitable module is
chosen automatically.
<visa-resource>: (Optional) Specify the address string of the N4877A.
Return Range
NONe|MUX|DMUX|BOTH
Description
This command controls the integration of a multiplexer and/or
demultiplexer to a data module. Also, it controls the integration of the
N4877A Clock Data Recovery and Demultiplexer.
In M8062A, the parameter values 'MUX', 'DMUX' and 'BOTH' all perform
the same action i.e. they enable the 32G mode. 'NONe' disables 32G
mode.
This SCPI is applicable for M8061A and M8062A.
Example
:CONF:MINT ‘M2.MuxMode’, BOTH