Keysight M8070A Programming Guide
167
SCPI Command Reference
5
[:SOURce]:JITTer:HFRequency:SPECtrally[:STATe][?]
Syntax
[:SOURce]:JITTer:HFRequency:SPECtrally[:STATe] ‘identifier’
<ON|OFF|1|0>
[:SOURce]:JITTer:HFRequency:SPECtrally[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', ‘M1.ClkOut’, or ‘M*.DataOut’
<ON|OFF|1|0>: Enable/disable spectrally distributed jitter.
Return Range
0|1
Description
This command enables/disables the generation of spectrally distributed
random jitter.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example
:JITT:HFR:SPEC ‘M2.DataOut2’, off
[:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*)[?]
Syntax
[:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*) ‘identifier’ <NRf>
[:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*)? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or ‘M*.DataOut’
<NRf>: Set spectrally distributed jitter amplitude.
Return Range
0 mUI to 78.71 mUI
Description
This command sets the spectrally distributed jitter amplitude in unit
intervals or time.
A suffix attached to the :AMPLitude(*) command is used to address two
jitter amplitudes. The :AMPL1 <NR3> command changes the level of the
internal RJLF resource and the :AMPL2 <NR3> command changes the
level of the internal RJHF resource. Both levels are specified in rms.
The units are set using the [:SOURce]:JITTer:HFRequency:UNIT command.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example
:JITT:HFR:SPEC:AMPL2 ‘M2.DataOut2’, 0.013