Keysight M8070A Programming Guide
313
SCPI Command Reference
5
:PLUGin:OTIMing:EVALuation:ETJBer[?]
Syntax
:PLUGin:OTIMing:EVALuation:ETJBer ‘identifier’, <EM6|...|EM12>
:PLUGin:OTIMing:EVALuation:ETJBer? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<EM6|...|EM12>: Specify the residual BER for total jitter estimation
calculation.
Return Range
EM6 to EM12
Description
This command specifies the residual BER for total jitter estimation
calculation and associates it with a specific measurement name identifier.
The range is 1E-12 (EM12) to 1E-6 (EM6). The query returns the current
residual BER for total jitter estimation calculation. The default is 1E-12.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example
:PLUG:OTIM:EVAL:ETJB ‘MyMeasurement’, EM7
:PLUGin:OTIMing:EVALuation:ERRType[?]
Syntax
:PLUGin:OTIMing:EVALuation:ERRType ‘identifier’, <AERR|ZERR|OERR>
:PLUGin:OTIMing:EVALuation:ERRType? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<AERR|ZERR|OERR>: Specify the error type to evaluate.
Return Range
AERR|ZERR|OERR
Description
This command specifies the error type to evaluate and associates it with a
specific measurement name identifier. AERR allows evaluation of all errors.
ZERR allows evaluation of errored zeroes and OERR allows evaluation of
errored ones. The query returns the current value.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example
:PLUG:OTIM:EVAL:ERRT ‘MyMeasurement’, ZERR