Keysight M8070A Programming Guide
309
SCPI Command Reference
5
The query returns the current location or group name identifier against
which the measurement name identifier is configured for the data
acquisition.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example
:PLUG:OTIM:ACQ:ALOC ‘MyMeasurement’,“M1.DataIn1”
:PLUGin:OTIMing:ACQuisition:DRESolution[?]
Syntax
:PLUGin:OTIMing:ACQuisition:DRESolution ‘identifier’, <NRf>
:PLUGin:OTIMing:ACQuisition:DRESolution? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the sample delay resolution.
Return Range
1 mUI to 250 mUI
Description
This command sets the sample delay resolution in terms of UI and
associates it with a specific measurement name identifier. The query
returns the current sample delay resolution value in UI.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example
:PLUG:OTIM:ACQ:DRES ‘MyMeasurement’, 0.25
:PLUGin:OTIMing:ACQuisition:FTJBer[?]
Syntax
:PLUGin:OTIMing:ACQuisition:FTJBer ‘identifier’, <NRf>
:PLUGin:OTIMing:ACQuisition:FTJBer? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Specify the BER value for the fast total jitter measurement.
Return Range
1E-15 to 1E-9
Description
This command specifies the BER value for the fast total jitter measurement
and associates it with a specific measurement name identifier. The query
returns the current value at which the fast total jitter measurement was
performed.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example
:PLUG:OTIM:ACQ:FTJB ‘MyMeasurement’, 1E-10