228
Keysight M8070A Programming Guide
5
SCPI Command Reference
:OUTPut:DEMBedding:SCABle:STATe[?]
Syntax
:OUTPut:DEMBedding[:SCABle][:STATe] ‘identifier’, <OFF | ON | 0 | 1>
:OUTPut:DEMBedding[:SCABle][:STATe]? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
OFF | ON | 0 | 1: Enables/disables standard cable compensation.
Return Range
0|1
Description
This command switches on/off the standard cable compensation on the
specified output.
The query return the present setting.
This SCPI is applicable for M8195A and M8196A.
Example
:OUTP:DEMB:SCAB:STAT ‘M1.DataOut1’, ON
:OUTPut:DEMBedding:SPARameter:STATe[?]
Syntax
:OUTPut:DEMBedding:SPARameter[:STATe] ‘identifier’, <OFF | ON | 0 | 1>
:OUTPut:DEMBedding:SPARameter[:STATe]? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
<OFF | ON | 0 | 1>
Description
This command switches on/off the compensation via S-parameter file on
the specified output.
The query return the present setting.
This SCPI is applicable for M8195A and M8196A.
Example
:OUTP:DEMB:SPAR:STAT ‘M1.DataOut1’, ON
:OUTPut:DEMBedding:SPARameter:PROFile[?]
Syntax
:OUTPut:DEMBedding:SPARameter:PROFile ‘identifier’, <”Filename”>
:OUTPut:DEMBedding:SPARameter:PROFile? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
Description
This command selects the S-Parameter profile. S-parameter is defined by
frequency/real/imaginary tuples for 1 to 9 port components.
The query return the present setting.