SCPI Command Reference
:PLUGin:ERATio:ACQuisition:AEND[?]
Syntax
:PLUGin:ERATio:ACQuisition:AEND ‘identifier’, <PFA|FDUR>
:PLUGin:ERATio:ACQuisition:AEND? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<PFA|FDUR>: Set accumulation end mode.
Return Range
PFA|FDUR
Description
This command sets the accumulation to end when a pass/fail (PFA)
condition has been met or end after the full duration (FDUR) of the
measurement for the addressed measurement name identifier. The query
returns the current accumulation end value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:ERAT:ACQ:AEND ‘MyMeasurement’, PFA