140
Keysight M8070A Programming Guide
5
SCPI Command Reference
Case 2
- The following description is applicable when the two channel
license has been purchased:
• SINGle: DataOut1 will be sourced from extended memory and
DataOut4 will be sourced from module internal memory. The data rate
range in this mode is 256 Mb/s … 32.5 Gb/s.
• DUAL: DataOut1 and DataOut4 will be sourced from extended memory.
The data rate range in this mode is 256 Mb/s … 16.250 Gb/s
• RANDom: All the available channels are used to generate random
interference.
• SINusoidal: All the available channels are used to generate random
interference.
Description
This command controls the memory configuration of M8195A AWG.
This query returns the current setting.
This SCPI is applicable for M8195A and M8196A.
Example
:MEM:CONF 'M1', FOUR
:MEM:CONF? 'M1'