408
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
27
MIL-STD-1553/ARINC 429 Triggering and Serial Decode
3
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
Word Start
— triggers on the start of a word.
•
Word Stop
— triggers at the end of a word.
•
Label
— triggers on the specified label value.
•
Label + Bits
— triggers on the label and the other word fields as specified.
•
Label Range
— triggers on a label following in a min/max range.
•
Parity Error
— triggers on words with a parity error.
•
Word Error
— triggers on an intra-word coding error.
•
Gap Error
— triggers on an inter-word gap error.
•
Word or Gap Error
— triggers on either a Word or Gap Error.
•
All Errors
— triggers on any of the above errors.
•
All Bits (Eye)
— triggers on any bit, which will therefore form an eye diagram.
•
All 0 Bits
— triggers on any bit with a value of zero.
•
All 1 Bits
— triggers on any bit with a value of one.
4
If you select the
Label
or
Label + Bits
condition, use the
Label
softkey to specify
the label value.
Label values are always displayed in octal.
5
If you select the
Label + Bits
condition, use the
Bits
softkey and submenu to
specify the bit values:
Use the
Define
softkey to select SDI, Data, or SSM. The SDI or SSM selections
may not be available, depending on word format selection in the Serial Decode
Menu.
Use the
Bit
softkey to select the bit to change.
Use the
0 1 X
softkey to set the bit value.
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...