Mask Testing
15
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
247
Region number 1 is the top mask region. The vertices in Region 1 describe points
along a line; that line is the bottom edge of the top portion of the mask.
Similarly, the vertices in Region 2 describe the line that forms the top of the
bottom part of the mask.
The vertices in a mask file are normalized. There are four parameters that define
how values are normalized:
• X1
•
Δ
X
• Y1
• Y2
These four parameters are defined in the Oscilloscope Setup portion of the mask
file.
The Y-values (normally voltage) are normalized in the file using the following
equation:
Y
norm
= (Y - Y1)/
Δ
Y
where
Δ
Y = Y2 - Y1
To convert the normalized Y-values in the mask file to voltage:
Y = (Y
norm
*
Δ
Y) + Y1
where
Δ
Y = Y2 - Y1
The X-values (normally time) are normalized in the file using the following
equation:
X
norm
= (X - X1)/
Δ
X
To convert the normalized X-values to time:
X = (X
norm
*
Δ
X) + X1
Oscilloscope
Setup Information
The keywords "setup" and "end_setup" (appearing alone on a line) define the
beginning and end of the oscilloscope setup region of the mask file. The
oscilloscope setup information contains remote programming language
commands that the oscilloscope executes when the mask file is loaded.
Any legal remote programming command can be entered in this section.
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...