242
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
15
Mask Testing
Run Until
The Run Until softkey lets you specify a condition on which to terminate testing.
• Forever
— The oscilloscope runs continuously. However, if an error occurs the
action specified using the
On Error
softkey will occur.
• Minimum # of Tests
— Choose this option and then use the
# of Tests
softkey
to select the number of times the oscilloscope will trigger, display the
waveform(s), and compare them to the mask. The oscilloscope will stop after
the specified number of tests have been completed. The specified minimum
number of tests may be exceeded. If an error occurs the action specified
using the
On Error
softkey will occur. The actual number of tests completed
is displayed above the softkeys.
• Minimum Time
— Choose this option and then use the
Test Time
softkey to
select how long the oscilloscope will run. When the selected time has passed
the oscilloscope will stop. The specified time may be exceeded. If an error
occurs the action specified using the
On Error
softkey will occur. The actual
test time is displayed above the softkeys.
• Minimum Sigma
— Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the oscilloscope
will perform the action specified by the
On Error
softkey.) Note that this is a
test sigma (the max achievable process sigma, assuming no defects, for a
certain number of tested waveforms) as opposed to a process sigma (which is
tied to the amount of failures per test). The sigma value may exceed the
selected value when a small sigma value is chosen. The actual sigma is
displayed.
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...