362
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24
FlexRay Triggering and Serial Decode
6
Press
Source
and select the analog channel that is probing the FlexRay signal.
7
Press
Threshold
; then, turn the Entry knob to set the threshold voltage level.
The threshold level should be set below the idle level.
The threshold voltage level is used in decoding and will become the trigger
level when the trigger type is set to the selected serial decode bus.
8
Press
Baud
and select the baud rate of the FlexRay signal being probed.
9
Press
Bus
and select the bus type of the FlexRay signal being probed.
It is important to specify the correct bus because this setting affects CRC error
detection.
10
Press
Auto Setup
to perform the following actions:
•
Sets the selected source channel's impedance to 50 Ohms, assuming a
differential active probe that requires a 50 ohm termination is being used .
•
Sets the selected source channel's probe attenuation to 10:1.
•
Sets the trigger level (on the selected source channel) to -300 mV.
•
Turns on trigger Noise Reject.
•
Turns on Serial Decode.
•
Sets the trigger type to FlexRay.
FlexRay Triggering
To set up the oscilloscope to capture a FlexRay signal, see
After you have set up the oscilloscope to capture a FlexRay signal, you can then
set up triggers on frames (see
), errors (see
), or events (see
).
NOTE
To display FlexRay serial decode, see
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...