Measurements
14
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
235
1
From the Measurement Menu, press the
Settings
softkey; then, press the
Thresholds
softkey to set analog channel measurement thresholds.
You can also open the Measurement Threshold Menu by pressing
[Analyze] >
Features
and then selecting
Measurement Thresholds
.
2
Press the
Source
softkey to select the analog channel or math waveform source
for which you want to change measurement thresholds.
Each analog channel and the math waveform can be assigned unique threshold
values.
3
Press the
Type
softkey to set the measurement threshold to
%
(percentage of
Top and Base value) or to
Absolute
(absolute value).
•
Percentage thresholds can be set from 5% to 95%.
•
The units for absolute threshold for each channel is set in the channel probe
menu.
•
When the
Source
is set to
Math: f(t)
, the threshold
Type
can only be set to
Percent
.
4
Press the
Lower
softkey; then, turn the Entry knob to set the lower measurement
threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from 5% to
93%.
TIP
Absolute threshold hints
•
Absolute thresholds are dependent on channel scaling, probe attenuation, and probe units. Always
set these values first before setting absolute thresholds.
•
The minimum and maximum threshold values are limited to on-screen values.
•
If any of the absolute threshold values are above or below the minimum or maximum waveform
values, the measurement may not be valid.
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...