Mask Testing
15
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
251
• The Y1 setup command specifies the vertical location of the x,y graph's y origin.
• The XDELta setup command specifies the amount of time associated with each
x unit.
• The Y2 setup command is the vertical location of the x,y graph's y=1 value (so
in effect, Y2 – Y1 is the YDELta value).
For example:
• With a graticule whose trigger position is 10 ns (before a center screen
reference) and whose ground reference (offset) is 2 V below the center of the
screen, to place the mask region's x,y graph's origin at center screen, you would
set X1 = 10 ns and Y1 = 2 V.
• If the XDELta parameter is set to 5 ns and Y2 is set to 4 V, a mask region whose
vertices are (-1, 1), (1, 1), (1, -1), and (-1, -1) goes from 5 ns to 15 ns and from
0 V to 4 V.
• If you move the mask region's x,y graph origin to the t=0 and V=0 location by
setting X1 = 0 and Y1 = 0, the same vertices define a region that goes from
-5 ns to 5 ns and from -2 V to 2 V.
How is mask testing done?
InfiniiVision oscilloscopes start mask testing by creating a database that is 200 x
640 for the waveform viewing area. Each location in the array is designated as
either a violation or a pass area. Each time a data point from a waveform occurs in
a violation area a failure is logged. If
Test All
was selected, every active analog
channel is tested against the mask database for each acquisition. Over 2 billion
failures can be logged per-channel. The number of acquisitions tested is also
logged and displayed as "# of Tests".
The mask file allows greater resolution than the 200 X 640 database. Some
quantization of data occurs to reduce the mask file data for display on-screen.
NOTE
Although a mask can have up to 8 regions, in any given vertical column, it is only possible to
define 4 regions. When there are 4 regions in a vertical column, one region must be tied to the
top (using the MAX y value) and one must be tied to the bottom (using the MIN y value).
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...