2- 16
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
Measurement Modes
4. After the settling detection stops, the quasi-pulse source keeps the output.
5. After the delay time, the measurement channel starts measurement.
Only one channel can be used for measurement.
6. After measurement, the quasi-pulse source immediately returns the output to the
start value and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE
If there is noise or skew on the output voltage, settling detection might stop at an
unexpected voltage.
NOTE
The BDT command sets the hold time and delay time, and the BDM command sets
the settling detection interval and measurement mode (voltage or current); the BDV
command sets the output. Also |start
-
stop| must be 10 V or more.