Keysight E5260/E5270 Programming Guide, Edition 4
2- 45
Remote Mode Functions
Initial Settings
Table 2-6
Initial Settings
Setup Item
Initial Setting
Commands
Measurement channel
Channel assigned the lowest number.
MCH
Measurement data to be displayed
Data 1
Compliance side data
MPA
Data 2
none
MPA
Output channel
Channel assigned the lowest number.
SCH
Output data to be displayed
Data 1
OUT
SPA
Data 2
CPL
SPA
Data display format
ENGINEERING
DFM
Remote mode data display
off
RED
Remote mode key lock/unlock
unlock
KLC
Auto calibration
on
CM
ADC zero function (for E5270B)
off
AZ
SMU output switch
open
CN, CL
ASU path/1 pA auto range/indicator
SMU side/disabled/enabled
SAP/SAR/SAL
Filter/Series resistor
off/off
FL/SSR
A/D converter
High speed ADC
AAD
Integration time (High resolution
ADC is available for E5270B)
High speed ADC: auto
AIT
High resolution ADC: auto
AIT
AV command parameter
number
=1,
mode
=0
AV
Current measurement range
with pulse
Compliance range
RI
without pulse auto
Voltage measurement range
with pulse
Compliance range
RV
without pulse auto
Sweep source parameters
cleared
WV, WSV, WI, WSI
Automatic abort function
off
WM
output after sweep measurement
Start value
WM
Pulse source parameters
cleared
PV, PI
Pulse sweep source parameters
cleared
PWV, PWI
Pulse width
0.001 s
PT
Pulse period
0.01 s
PT
Search source parameters
cleared
BSV, BSSV, BSI, BSSI, LSV, LSSV, LSI, LSSV
Search monitor parameters
cleared
BGV, BGI, LGV, LGI
Output after search measurement
Start value
BSM, LSM
Search measurement data
Source output value only
BSVM, LSVM
Quasi-pulse source parameters
cleared
BDV
Quasi-pulsed spot measurement mode Voltage
BDM
Quasi-pulse settling detection interval Short
BDM
Hold time
0 s
WT, PT, BDT, BST, LSTM
Delay time
0 s
WT, PT, BDT, BST, LSTM
Step delay time
0 s
WT
Trigger delay time
0 s
WT, PT
Trigger mode
XE, TV, TI, or GET
TM
Trigger port
Ext Trig In
Start Measurement trigger input
TGP
Ext Trig Out
Measurement Completion trigger output TGP
Digital I/O
cleared
TGP
Trigger condition of
Start Step Output Setup trigger
with trigger for first sweep step
TGSI
Type of output trigger
Edge trigger
TGXO, TGSO, TGMO
Digital I/O port
Output for all port
ERM
Program memory
cleared. Not cleared by *RST command or device clear.
SCR
Value of internal variable (%In, %Rn) 0
VAR
Data output format
ASCII with header, CR/LF^EOI
FMT
Data output buffer
cleared
BC
Status byte
Only bit 6 is enabled.
*SRE
Error code register
cleared
ERR?
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...