1- 30
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Data Output Format
Binary Data Output Format
This section describes the binary data output format, and the elements of the data.
•
•
NOTE
Data resolution
The resolution of binary data is as shown below.
•
Measurement data: Measurement range / 50000
•
Output data: Output range / 20000
Note that the resolution of the measurement data is larger than the resolution of the
high resolution A/D converter.
Data Format
The data output format depends on the measurement mode as shown below.
High Speed Spot
Data
<terminator>
Data
is the value measured by the channel you specify by using the TI or TV
command.
Spot
Data1
[
Data2
] . . . . <terminator>
DataN
(
N
: integer) is the value measured by a channel. The order of
Data
is defined
by the MM command.
Pulsed Spot,
Quasi-Pulsed Spot
Data
<terminator>
Data
is the value measured by the channel you specify by using the MM command.
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......