4- 150
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
XE
The XE command triggers the E5260/E5270 to start measurement, or causes the
E5260/E5270 to recover from the wait state set by the PA command.
This command is not available for the high-speed spot measurement.
NOTE
After measurement, the measurement data will be entered to the output data buffer.
For data output format, refer to
“Data Output Format” on page 1-22
.
Execution
Conditions
The following execution conditions are for you who use the XE command to start
measurement. There is no execution condition when you use the XE command to
recover from the wait state.
•
If any channel is set to the HIGH VOLTAGE state (forcing more than ±42 V, or
voltage compliance set to more than ±42 V) after the trigger (XE), the interlock
terminal must be shorted.
•
The following commands must be entered before the XE command.
Syntax
XE
Example
Statement
OUTPUT @E5270;"XE"
Measurement Mode
Commands
Spot
CN, MM, DV or DI
Staircase sweep
CN, MM, WV or WI
Pulsed spot
CN, MM, PV or PI
Pulsed sweep
CN, MM, PWV or PWI
Staircase sweep with pulsed bias
CN, MM, WV or WI, PV or PI
Quasi-pulsed spot
CN, MM, BDV
Liner search
CN, MM, LSV or LSI, LGV or LGI
Binary search
CN, MM, BSV or BSI, BGV or BGI
Multi channel sweep
CN, MM, WI or WV, WNX
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......