Keysight E5260/E5270 Programming Guide, Edition 4
2- 11
Remote Mode Functions
Measurement Modes
NOTE
The PT command sets the hold time, pulse width, and pulse period. The WM
command sets the automatic abort function and the output after measurement. The
PWV sets the pulsed sweep voltage, and the PWI sets the pulsed sweep current. The
start, stop, and pulse base values must have the same polarity for log sweep.
To Use
Synchronous
Sweep Source
One more channel can be set up as a staircase sweep source that has the output
synchronized with the pulsed sweep. Refer to
“Synchronous Output” on page 2-21
.
After the measurement, the synchronous sweep source forces the start value, and
keeps it.
Figure 2-7
Synchronous Sweep
NOTE
The WSV command sets the sweep voltage, and the WSI command sets the sweep
current. You can use the same output mode (voltage or current) as the pulsed sweep.
The start and stop values must have the same polarity for log sweep.
To Stop Sweep
Output
An automatic abort function is available. Refer to
Even if the automatic abort function is disabled, the E5260/E5270 automatically
stops measurement if power compliance is enabled for the sweep source and the
power compliance or an automatic abort condition is detected.
Voltage or current
Start value
Time
Stop value
Pulse base value
Stop value
Synchronous sweep
Pulsed sweep
Previous value
Trigger (e.g. XE)
PT/WM/PWV/PWI
WSV/WSI
Start value
Previous value
Voltage or current
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......