Keysight E5260/E5270 Programming Guide, Edition 4
3- 15
Programming Examples
Pulsed Spot Measurements
Pulsed Spot Measurements
To perform pulsed spot measurements, use the following commands.
NOTE
Measurement channel performs measurement so that the pulse width and pulse
period are kept. The integration time is automatically set by the instrument, and you
cannot change. For the Keysight E5270B, note that the high-resolution ADC cannot
be used for the pulsed measurements. The AAD/AIT/AV/WT command settings are
ignored.
Function
Command
Parameters
Enables channels
CN
[
chnum
... [,
chnum
] ... ]
Disables channels
CL
[
chnum
... [,
chnum
] ... ]
Sets filter ON/OFF
[FL]
mode
[,
chnum
... [,
chnum
] ... ]
Sets series resistor ON/OFF
[SSR]
chnum
[,
mode
]
Forces constant voltage
DV, TDV
chnum
,
vrange
,
output
[,
comp
[,
polarity
[,
irange
]]]
Forces constant current
DI, TDI
chnum
,
irange
,
output
[,
comp
[,
polarity
[,
vrange
]]]
Sets pulse timing parameters
PT
hold
,
width
[,
period
[,
tdelay
]]
Forces pulse voltage
PV
chnum
,
range
,
base
,
pulse
[,
comp
]
Forces pulse current
PI
chnum
,
range
,
base
,
pulse
[,
comp
]
Sets voltage measurement
range
[RV]
chnum
,
range
Sets current measurement
range
[RI]
chnum
,
range
[RM]
chnum
,
mode
[,
rate
]
Selects measurement mode
MM
3,
chnum
Sets SMU operation mode
[CMM]
chnum
,
mode
Executes measurement
XE
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......