Command Parameters
Table 4-4
Voltage Output Ranging Type
Table 4-5
Current Output Ranging Type
range
or
vrange
Ranging type
0
Auto ranging
5
0.5 V limited auto ranging, for E5281B/E5287A
50
5 V limited auto ranging, for E5281B/E5287A
20 or 11
2 V limited auto ranging
200 or 12
20 V limited auto ranging
400 or 13
40 V limited auto ranging
1000 or 14
100 V limited auto ranging
2000 or 15
200 V limited auto ranging, for E5280B/E5290A
range
or
irange
Ranging type
0
Auto ranging
8
1 pA limited auto ranging, for E5288A, not available for pulsed output
9
10 pA limited auto ranging, for E5287A, not available for pulsed output
10
100 pA limited auto ranging, for E5287A, not available for pulsed output
11
1 nA limited auto ranging, for E5280B/E5281B/E5287A, not available for pulsed output
12
10 nA limited auto ranging, for E5280B/E5281B/E5287A
13
100 nA limited auto ranging
14
1
μ
A limited auto ranging
15
10
μ
A limited auto ranging
16
100
μ
A limited auto ranging
17
1 mA limited auto ranging
18
10 mA limited auto ranging
19
100 mA limited auto ranging
20
200 mA limited auto ranging, for E5291A
1 A limited auto ranging, for E5280B/E5290A
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......