Keysight E5260/E5270 Programming Guide, Edition 4
4- 121
Command Reference
Command Reference
TSC
The TSC command enables or disables the time stamp function.
This command is effective for ASCII data output format. Refer to
.
Execution
Conditions
Time stamp function is not available for the following measurement modes:
•
Quasi-pulsed spot measurement (MM 9)
•
Linear search measurement (MM 14)
•
Binary search measurement (MM 15)
Syntax
TSC
mode
Parameters
mode
:
Time stamp function mode. Integer expression.
When the function is enabled, the E5260/E5270 returns the time data
with the measurement data. The time data is the time from when the
timer is cleared until the measurement is started. Refer to
.
Remarks
To read the time data with the best resolution (100
μ
s), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.
Example
Statements
OUTPUT @E5270;"TSC 1"
mode
Description
0
Disables the time stamp function. Initial setting.
1
Enables the time stamp function.
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......