4- 86
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
OS
The OS command causes the E5260/E5270 to send a edge trigger from the Ext Trig
Out terminal. To set the trigger logic (initial value: negative), send the TGP
command for the Ext Trig Out terminal.
Syntax
OS
Example
Statements
OUTPUT @E5270;"OS"
OSX
The OSX command causes the E5260/E5270 to send a trigger from a trigger output
terminal specified by the
port
parameter. To set the trigger logic (initial value:
negative), send the TGP command for the specified port.
Syntax
OSX
port
[,
level
]
Parameters
port
:
External trigger output port number. Integer expression.
-
2, or 1 to 16.
-
2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
To use a digital I/O port, send the TGP command. The
port
value must
be same as the
port
value set to the TGP command.
level
:
Trigger output level. Integer expression. 0, 1, or 2.
0: Logical low.
1: Logical high.
2: Edge trigger (default setting).
If
level
is not specified, the E5260/E5270 sends the edge trigger. For
the gate trigger output, send OSX
port
,1 when starting trigger output,
and send OSX
port
,0 when stopping trigger output.
Example
Statements
OUTPUT @E5270;"OSX 1,1"
OUTPUT @E5270;"TI";1
ENTER @E5270 USING "#,3X,12D,X";Idata
OUTPUT @E5270;"OSX 1,0"
See Also
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......