![Keysight Technologies E5260 Series Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/keysight-technologies/e5260-series/e5260-series_programming-manual_1961978157.webp)
Keysight E5260/E5270 Programming Guide, Edition 4
3- 49
Programming Examples
Multi Channel Sweep Measurements
Measurement
Result Example
Vb (V), Ib (mA), Tb (sec), Stat_b, Ic (mA), Tc (sec), Stat_c
0.1, 0.01, 0.02949, NDI, -0.025, 0.02949, NCI
0.18, 0.01, 0.03788, NDI, -0.03, 0.03788, NCI
0.26, 0.01, 0.04628, NDI, -0.03, 0.04628, NCI
0.34, 0.01, 0.05468, NDI, -0.025, 0.05468, NCI
0.42, 0.01, 0.06308, NDI, -0.025, 0.06308, NCI
0.5, 0.02, 0.07148, NDI, -0.025, 0.07148, NCI
0.58, 0.105, 0.07987, NDI, 0.005, 0.07987, NCI
0.66, 0.585, 0.08825, NDI, 0.5, 0.08825, NCI
0.74, 2.635, 0.09664, NDI, 4.885, 0.09664, NCI
0.82, 9.96, 0.10505, NDI, 20.5, 0.10505, NCI
0.9, 27.84, 0.11345, NDI, 45.75, 0.11345, NCI
Data save completed.
Do you want to perform measurement again?
For i = 0 To nop1 - 1 ’58
tm1(i) = Val(Mid(mret, 4 + 16 * 5 * i, 12))
st1(i) = Mid(mret, 17 + 16 * 5 * i, 3)
md1(i) = Val(Mid(mret, 20 + 16 * 5 * i, 12))
tm2(i) = Val(Mid(mret, 36 + 16 * 5 * i, 12))
st2(i) = Mid(mret, 49 + 16 * 5 * i, 3)
md2(i) = Val(Mid(mret, 52 + 16 * 5 * i, 12))
sc(i) = Val(Mid(mret, 68 + 16 * 5 * i, 12))
data(j, i) = Chr(13) & Chr(10) & sc(i) & ", " & md1(i) * 1000 & ", " & tm1(i) &
", " & st1(i) & ", " & md2(i) * 1000 & ", " & tm2(i) & ", " & st2(i)
Next
session.WriteString("DZ" & vbLf) ’69
save_data(fname, title, value, data, nop1, nop2, session, t)
Exit Sub
Check_err: ’73
session.WriteString("EMG? " & err & vbLf) : msg = session.ReadString(256)
MsgBox("Instrument error: " & err & Chr(10) & msg, vbOKOnly, "")
Exit Sub
Check_nop:
MsgBox("No. of data: " & rep & " (not " & nop1 * 5 & ")", vbOKOnly, "") ’79
End Sub
Line
Description
58 to 67
Picks the measurement data out and stores it into the
data
array.
69 to 71
Applies 0 V from all channels and transfers the data stored in the
data
variable to the
save_data subprogram (see
). And the subprogram will save the data into a
CSV file specified by the
fname
variable and displays the data on a message box.
73 to 79
Displays a message box to show an error message if the error is detected. Also displays
a message box to show an error message if the number of returned data is not correct.
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......