Keysight E5260/E5270 Programming Guide, Edition 4
5- 5
Error Messages
Error Codes
126
Pulse base and peak must be same polarity.
The polarity of the base and peak values must be the same in the PI
command. Also the polarity of the base, start, and stop values must be
the same in the PWI command.
130
Start and stop must be same polarity.
For a log sweep, the polarity of the start and stop values must be the
same in the WV, WI, WSV, WSI, or WNX command. Also, 0 is not
allowed for the start and stop values.
150
Command input buffer is full.
The Keysight E5260/E5270 can receive 256 characters maximum
including the terminator at one time.
152
Cannot use failed module.
The channel number specifying the module failed the self-test or
calibration. Specify another module that passed the self-test or
calibration. For the service purpose, execute the RCV command to
enable the module.
153
No module for the specified channel.
Module is not installed in the slot specified by the channel number.
160
Incorrect ST execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not enter the ST command
between the ST command and the END command.
161
Incorrect END execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not send the END command
before starting the programming.
162
Incorrect command for program memory.
Specified command cannot be stored in the program memory. For the
incorrect commands, refer to
.
170
Incorrect usage of internal variable.
The internal variable must be %I
n
for integer data, or %R
n
for real
data. where
n
is an integer, 0 to 99. Use %I
n
for the integer type
command parameters; and use %R
n
for the real type command
parameters. For the internal variables, refer to
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......