4- 148
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
step
:
Number of steps for staircase sweep. Numeric expression. 1 to 1001.
The E5260/E5270 can store approximately 16000 measurement data in
ASCII format with header without source data, or approximately 64000
measurement data in binary format.
Icomp
:
Current compliance (in A). Numeric expression. See
. If you do not set
Icomp
, the previous value is used.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified
Icomp
. If the output value is 0,
the compliance polarity is positive.
If you set
Pcomp
, the maximum
Icomp
value for the module is allowed,
regardless of the output range setting.
Pcomp
:
Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the
Pcomp
value is not entered, the power compliance is not set.
0.001 to 2, 0.001 to 4 for E5291A, or 0.001 to 20 for E5280B/E5290A
Example
Statements
OUTPUT @E5270;"WV 1,2,12,1E-6,10,100,0.1,1"
OUTPUT @E5270;"WV 2,1,0,0,20,101"
Содержание E5260 Series
Страница 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Страница 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Страница 13: ...1 Programming Basics...
Страница 63: ...2 Remote Mode Functions...
Страница 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Страница 109: ...3 Programming Examples...
Страница 185: ...4 Command Reference...
Страница 335: ...5 Error Messages...
Страница 349: ......