1. Preparations for measurement
MKC-710 9
1-3-2.Installation of Auto Solvent Change Unit
1)
Fix the bottle for dehydrated solvent and bottle for waste with the cap (waste bottle
cover) respectively.
ポリ瓶
脱水溶剤びん
Note
The bottle caps must be securely fixed in order to avoid air leak, which would
prevent the dispenser from working properly in suction and draining.
2)
Connect the silicone tubes on both ends of desiccant tube, and put the tube in place as
shown below
3)
Put the above 1) bottle in reagent bottle holder. If the outside diameter of solvent
bottle does not match the holder, use the bottle holder and stand as shown below.
4)
Connect the silicon tubes, one to the top of solvent bottle.
Waste bottle
Bottle for dehydrated solvent
Silicone tube