6. Setup
MKC-710 83
Other
6-9.
Display/printout format and automatic setting of mean values can be set up.
Select “
9
.Other” with [
↑
][
↓
] and press [ENTER].
[
Character Display (Font size)
]
Select the font size of measurement results on screen. Select with [
←
] [
→
] keys,
and press [ENTER].
x
Normal : Results are shown in normal size.
x
Large :
Results are shown in twice the size of “Normal.”
[Print Header]
Make setting of header
printing. The header includes ‘Model name’, ‘Serial No.’ and
‘Print date’.
Select with [
←
] [
→
] keys, and press [ENTER].
x
Off
:
No header is printed.
x
On
:
Header is printed.
[Print Footer]
Make setting of footer printing. The footer includes a printing operator. Select with
[
←
] [
→
] keys, and press [ENTER].
x
Off
:
No footer is printed.
x
On
:
Footer is printed.
[Auto Set., mean]
The average value of a plural number of blank levels that have been measured will
be automatically set into the blank value to be used in sample setup to be used in
reagent information respectively.
Select “Auto Set. mean” with [
←
] [
→
] keys, and
press [ENTER].
x
Off
:
No setting. Each measurement result is put in the blank or factor.
x
On
:
Auto set in the blank. A mean value of up to five (5) results is obtained.
The value for “Auto Set. mean” is cleared when the method No. is
changed on sample setting or when the titrator is turned off.