8. Troubleshooting
98
Drift level is too high
8-3.
START
Are joints greased?
Is septum new?
Repair by
manufacturer
END
Apply grease
Change with new one
NO
NO
YES
Is reagent correct?
YES
Use correct reagent
NO
YES
Is reagent new?
Change with new one
NO
YES
Is inner burette dry?
Dry the burette
NO
Is anode in contact
with diaphragm?
Use anode adjuster to detach
from each other 1.5~2.0mm
NO
YES
YES
Inner burette or
Electrode good?
Change it if malfunctioning
NO
YES