4. Method
42
[Control
Gain]
Here you enter a coefficient for electrolysis speed, typically 5.0. If reaction speed is
so fast that an over-titration is expected, decrease the value. To the contrary, if
reaction speed is expected to be slow with a sample which contains a large amount
of water, increase this value.
1.0
~
9.9
[Electrolysis speed]
Selection of electrolysis mode:
Standard
:
Use this mode for regular type of measurement.
Fast
:
Use this mode for a sample with water more than 5000ug/H2O.
Note
Measurement results may be deviated with
“
Fast
”
mode more than
“
Standard
”
.
Check on the expected precision with the sample you are going to measure.
[Stable]
To display “Stable” status, you need to enter the drift change rate
down lower than preset level. Select a larger amount for the rate in case the
moisture in carrier gas is unstable to start titration. However, this means it will
increase the error and deviation of measurement results.
0.00
~
99.99
μ
g/min