9. Others
MKC-710 119
9-6-2.Method parameters
[Parameter and default]
Method No.
01
02
03
04
05
06
07-20
[Titration Parameter]
Titration Mode
H2O
H2O
H2O
H2O
Br2
H2O
H2O
Cell Type
2-Comp.
2-Comp.
2-Comp.
2-Comp.
2-Comp.
2-Comp.
2-Comp.
t(stir)
0s
0s
0s
0s
0s
0s
0s
t(wait)
15s
15s
15s
15s
15s
15s
15s
t(max)
0s
0
s
1200s
1200s
0s
0s
0s
Drift Stop
Rel.
Rel.
Off
Off
Rel.
Rel.
Rel.
Rel.
0.10ug/s
0.10ug/s
-
-
1.00ug/s
0.10ug/s
0.10ug/s
Abs.
-
-
-
-
-
-
-
Control Gain
5.0
5.0
5.0
5.0
5.0
5.0
5.0
Electrolysis Speed Standard
Standard
Standard
Standard
Standard
Standard
Standard
Stable
0.1ug/
min
0.1ug/
min
0.1ug/
min
0.1ug/
min
0.5ug/
min
0.1ug/
min
0.1ug/
min
Start Mode
Manual
Manual
Auto
Auto
Manual
Manual
Manual
Data Sampling
Time
5s
5s
5s
5s
5s
5s
5s
[Caluculation Parameter]
Caluculation Type
Sample
Sample
Sample
Sample
Sample
Check
Sample
Caluculation No.
1
2
2
2
7
2
1
Unit
ug
ppm
ppm
ppm
mg/100g
ppm
ug
Weight Input
Variable
Variable
Variable
Variable
Variable
Variable
Variable
Drift Comp.
Auto
Auto
Auto
Auto
Auto
Auto
Auto
Drift
-
-
-
-
-
-
-
Standard Value
-
-
-
-
-
0.0000
-
Permit. Error
-
-
-
-
-
0.0000
-
[Report]
Format
Short
Short
Short
Short
Short
Short
Short
Data List
Off
Off
Off
Off
Off
Off
Off
Graph
Off
Off
Off
Off
Off
Off
Off