8. Troubleshooting
MKC-710 99
It runs into over-titration
8-4.
START
Away from sunlight?
Is reagent new?
Repair by
manufacturer
END
Keep away from sunlight
Change with new one
NO
NO
YES
YES
YES
Is reagent correct?
Use correct reagent
NO
YES
YES
Is electrode
working alright?
Clean the electrode
NO
Change with
correct rotor (35mm)
Make stirrer speed increase
Is the length of
stirrer rotor normal?
Is stirrer speed appropriate?
YES
NO
NO