3. Basic operation
MKC-710 25
To change the life value of anolyte to zero, press [Reagent]
→
[Anode]
→
[Life]
→
current life value
→
[0]to[ENTER]key
<
When use the Auto Solvent Change Unit
>
<
injection
>
1)
Press [MENU/HOME] key.
Select “1.Manual Operation” and press
[ENTER] key. Select [Pump].
Press Injection Pump [On] key to
inject approx. 100mL of anolyte into
the titration cell. (The lower line of the
titration cell is a 100mL mark.)
[Off
→
On] key automatically changes
to [On
→
Off] at this time.
Press Injection Pump again, and
injection will stop.
2)
After changing anolyte, set the life value of anolyte to 0.