5. Function Tools
MKC-710 53
[Pump]
When connecting with the auto dispensing system, you can turn on or off the injection
pump and the drain pump. Refer to 3-3-2. Anolyte <When use the Auto Solvent Change
Unit> for details.
5-1-2.
Data File
Press [MENU/HOME], select
“2.Data File”
with [
↑
][
↓
] and press [ENTER]
[ Display of Titration result and Recalculation]
Move the cursor to a result you wish to display and press [ENTER] to display a result.
The result of After Titr is displayed respectively. Select the result and press [ENTER].
The results can be re-calculated.
Move the cursor to [Print] and press [ENTER] to reprint and recalculate the result. The
results of recalculation are printed out with sample number (Sample No.) headed with
(#) mark.
Note
Up to 100 samples measurement results can be stored. As for After Titr. result,
2 results are stored.
When it exceeds 100, note that data will be erased on the first-in first-out basis.
[Re-Calculation]
Move the cursor to [Re-Calculation] and press [ENTER] to recalculate. Sample size
and unit of the result and print format can be changed.
x
Lot
:
Change a LotNo.
x
Blank
:
Change a blank value
x
FA
:
Change a factor value
Report
x
Format
:
Change a print format of result
x
Data List
:
Change a print format of data list
x
Graph
:
Change a print format of titration curve
[Execute]
Execute to recalculate. A result screen is displayed.
[Exit]
Cancel to recalculate. A result screen is displayed without recalculation.