7. Maintenance
MKC-710 91
7-2-3.Cleaning the titration cell
1) Remove Detection electrode, Inner burette and Sampling port stopper, and
then drain out the reagent.
2) Wipe off the grease around the sliding area with methanol.
3) Rinse by neutral detergent under running water.
4) After drying the glassware in a heater, either cool them in desiccator or dry
them.
Note
Insufficiently dried glassware may cause higher drift level.
7-2-4.Distance adjustment between anode electrode and diaphragm
If the anode electrode in the inner burette and the diaphragm are too
close together, electrolysis reaction will not run in normal condition.
Use the supplied anode adjuster to adjust the distance in between.