1. Preparations for measurement
12
The following chloroform-containing reagents (see table below) or oil-based
samples deteriorate the drain pump.
<Do not use the standard drain pump with the following Anolyte:>
KEM
Fluka
(
RdH
)
& HPC
Mitsubishi
Anolyte
AO
AK
AS
Therefore, when using such reagents or samples, change the connecting of the Auto
Solvent Change Unit as follows.
<Installation (in the negative pressure)>
Connect the Drain tube to Cell to waste bottle. (Use Stopper and Long Male Union
supplied to Drain tube to Waste Bottle.
Connect the Suction tube (Bottle Cap
–
Pump) to waste bottle.
Connect the Suction tube (Bottle Cap
–
Pump) to magnetic stirrer back side pump IN.
Note
Please refer to a previous page for the connection of other parts.
Injection tube
Suction Tube
(Bottle Cap- Pump)
Drain tube to Cell
Waste bottle
Bottle for
Anolyte
Long Male Union
Stopper