6482-901-01 Rev. A / August 2012
Return to
8-11
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 8: Triggering
Figure 8-7
Operation model for triggering example
A) Turning the Model 6482 output on places it at point A in the flow chart, where it waits for an
external trigger.
B) Pressing the
STEP
key takes the Model 7001/7002 out of the idle state and places operation at
point B in the flow chart.
C) For the first pass through the model, the scanner does not wait at point B. Instead, it closes the
first channel (point C).
D) After the relay settles, the Model 7001/7002 outputs a trigger pulse. Because the instrument is
programmed to scan 10 channels, operation loops back to point B, where it waits for an input
trigger.
E) and F) With the Model 6482 operation at point A, the output trigger pulse from the Model 7001/
7002 triggers a measurement of device under test (DUT) #1 (point E). After the measurement is
complete, the Model 6482 outputs a trigger pulse and then loops back to point A, where it waits for
another input trigger.
The trigger applied to the Model 7001/7002 from the Model 6482 closes the next channel in the
scan, which then triggers the Model 6482 to measure that device under test (DUT). This process
continues until all 10 channels are scanned and measured.
Idle
Bypass
Scan
channel
Trigger
No
Yes
B
C
D
Idle
Make
measurement
Trigger
No
Yes
Model 6482
A
E
F
Made
10
measurements
?
Scanned
10
channels
?
7001 or 7002
Press STEP to start scan
Wait for
trigger link
trigger
Wait for
trigger link
trigger
Output
trigger
Output
trigger