16-26
Return to
6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
ENABle <b>
:DISPlay:ENABle <b>
Control display circuitry
Parameters
<b> =
0 or OFF
Disable display circuitry
1 or ON
Enable display circuitry
Query
:ENABle?
Query state of display
Description
This command is used to enable and disable the front panel display circuitry.
When disabled, the instrument operates at a higher speed. While disabled, the
display is frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL and OUTPUT
OFF) are disabled. Normal display operation can be resumed by using the
:ENABle command to enable the display or by putting the Model 6482 into local.
MODE <name>
:DISPlay:MODE <name>
Select CALC block for display
Parameters
<name> = CALCulate3
Display CALC3 (Channel 1 REL)
CALCulate4
Display CALC4 (Channel 2 REL)
CALCulate5
Display CALC5 (RATIO)
CALCulate6
Display CALC6 (DELTA)
DUAL
Dual-channel display mode
Query
:MODE?
Query selected display CALC block
Description
This command is used to select which CALC block to display. Displayed CALC
blocks include CALC3 (channel 1 REL), CALC4 (channel 2 REL), CALC5
(RATIO), and CALC6 (DELTA). The DUAL parameter selects the dual-channel
display mode.
ATTRibutes?
:DISPlay[:WINDow[1]]:ATTRibutes?Query attributes; top display
:DISPlay:WINDow2:ATTRibutes?Query attributes; bottom display
Description
This query command is used to determine which characters on the display are
blinking and which are not. The response message provides that status of each
character position for the specified display. The primary display consists of 20
characters and the secondary display consists of 32 characters.
1 = Character is blinking
0 = Character is not blinking
For example, assume the following menu is displayed with the SAVESETUP
option blinking:
MAIN MENU
SAVESETUP COMMUNICATION CAL >
The response message for :disp:attr? (top display) will result in 20 zeroes as fol-
lows:
00000000000000000000
The response message for :disp:wind2:attr? (bottom display) will display ones at
the character positions for SAVESETUP as follows:
11111111100000000000000000000000
Содержание 6482
Страница 3: ......
Страница 62: ...Section 4 Range Digits Speed and Filters In this section Topic Page Range and digits 4 2 Speed 4 4 Filters 4 6...
Страница 90: ...Section 7 Sweep Operation In this section Topic Page Sweep types 7 2 Configuring and running a sweep 7 5...
Страница 182: ...Section 14 Common Commands In this section Topic Page Command summary 14 2 Command reference 14 2...