Connection precautions
The following paragraphs summarize precautions that should be observed when making test
connections to the Model 6482.
WARNING
Maximum common-mode voltage (voltage between INPUT analog common
and chassis ground) is 200 V. Exceeding this value may result in a shock
hazard.
When making connections, do not leave any exposed connections. Be sure
that all external circuits are properly insulated.
CAUTION
Maximum INPUT and OUTPUT connector rating is 30 V at 20 mA. Exceeding
these values may result in instrument damage.
Input and output connectors
shows the location of the input and output connectors on the rear panel. These
connectors have the following functions.
•
Input connectors
INPUT CHANNEL 1 and CHANNEL 2: A 3-lug triaxial connector for direct current input from
the device under test (DUT) to channel 1 (or channel 2).
•
Analog output connectors
ANALOG OUT CHANNEL 1 and CHANNEL 2: A 3-lug triaxial connector that provides a
-10 V to +10 V DC output analogous to the input current applied to channel 1 or channel 2.
•
Voltage source (output) connectors
VOLTAGE SOURCE OUT CHANNEL 1 or CHANNEL 2: A safety banana jack is used to
apply the channel 1 HI or channel 2 HI voltage bias source signal to the DUT.
Figure 2-1
Model 6482 rear panel showing input and output connectors
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
SERIAL NO.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT
I
MADE IN
U.S.A.
OUTPUT ENABLE
RS-232
TRIGGER LINK
INPUT
RATINGS MAX.
30 V @ 20mA
CHANNEL 1
CHANNEL 2
RATINGS MAX.
COMMON
MODE
200V
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
ANALOG
OUT
RATINGS MAX.
30 V @ 20mA
VOLTAGE
SOURCE
OUT
30 V @ 20mA
Input connectors
Voltage
source (output)
connectors
Analog output
connectors
Model 6482