7-8
Return to
6482-901-01 Rev. A / August 2012
Section 7: Sweep Operation
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
NOTE
The Model 6482 limits the number of sweeps (sweep-count) according to the number of
steps required for each sweep. The maximum number of finite sweeps that can be
performed is determined as follows: Maximum finite sweep count = 3000 / # points in
sweep.
7. Choose
FINITE
or
INFINITE
.
8. From the CONFIGURE SWEEPS menu, select
SOURCE-RANGING
, press the
ENTER
key,
and then select
BEST-FIXED
,
AUTO-RANGE
, or
FIXED
.
9. Press the
EXIT
key to return to the normal display.
Step 3. Set the delay.
Set the source delay as follows:
1. Press the
CONFIG
key, and then the
SRC1
or
SRC2
key depending on the selected source
channel.
2. Select
DELAY
, and then press the
ENTER
key.
3. Set the delay value, and then press the
ENTER
key.
Step 4. Turn the output on.
Press the
ON/OFF OUTPUT
key to turn the output on (blue OUTPUT indicator turns on). The
Model 6482 will output the programmed bias level.
Step 5. Run the sweep.
To run the sweep, press the
SWEEP
key. After the sweep is completed, turn the output off by
pressing the
ON/OFF OUTPUT
key.
Step 6. Read the buffer.
Press the
RECALL
key to access the readings stored in the buffer. Press the
DISPLAY TOGGLE
key to display statistical information. For additional information on statistical information, see
Performing a logarithmic staircase sweep
Step 1. Configure the channel.
Configure the Model 6482 for the needed operations as follows:
1. Select the source channel by pressing the
SRC1
or
SCR2
key.
2. Set the source level.
3. Press
MSR1
or
MSR2
key to select the measurement channel, and then choose the
measurement range.
Step 2. Configure the sweep.
Configure the sweep as follows:
1. Press the
SRC1
or
SRC2
key to select the channel, and then press the
EXIT
key to leave edit
mode.
2. Press the
CONFIG
key and then the
SWEEP
key.
3. Select
TYPE
, and then press the
ENTER
key.
4. Select
LOG
, and then press the
ENTER
key.
5. At the prompts, enter values for
START
,
STOP
, and
NO-OF-POINTS
.
Содержание 6482