6482-901-01 Rev. A / August 2012
Return to
17-27
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 17: Calibration
10. Press the
ENTER
key. The instrument will prompt for the zero current:
I (0)
2
= +0.000000 nA
Use
,
,
,
, ENTER or EXIT.
11. Set the Model 6430 current to 0 nA and press the
ENTER
key. The instrument will prompt for
the negative full-range current:
I(-FS)
2
= -2.000000 nA
Use
,
,
,
, ENTER or EXIT.
12. Set the Model 6430 current to -1.90000 nA.
13. Adjust the displayed current to match the Model 6430 current, and then press the
ENTER
key
to complete calibration of the present range.
14. Press the
EXIT
key to return to normal display.
15. Repeat steps 3 through 14 for the 20 nA to 2
μ
A ranges, using
as a guide. Be sure
to set the Model 6482 to the correct range using the RANGE
and
keys, and use the
corresponding Model 6430 currents for each range.
16. Connect the Model 5700A to the Model 6482 channel 1 INPUT connector, as shown in
.
17. Select the Model 5700A V/R amperes function and set the Model 5700A output to
+19.0000
μ
A, and then turn on the output.
18. Press the Model 6482
MSR2
key, and then set channel 1 to the 20
μ
A current range using the
RANGE
and
keys (the 20
μ
A range has the following format: +00.0000
μ
A).
19. Press the
MENU
key, select
CAL
, and then press the
ENTER
key.
20. Select
EXECUTE
, and then press the
ENTER
key to enter the CAL EXECUTION menu.
21. Select
I-CAL
and press the
ENTER
key. The instrument will display the following:
I-CAL CHANNEL SELECT
CHAN-1 CHAN-2
22. Select
CHAN-2
and press the
ENTER
key. The instrument will prompt for the positive full-range
current:
I (+FS)
2
= +20.00000 uA
Use
,
,
,
, ENTER or EXIT.
23. Set the Model 5700A current to +19.0000
μ
A.
24. Use the EDIT keys to adjust the Model 6482 display value to match the Model 5700A current.
25. Press the
ENTER
key. The instrument will prompt for the zero current:
I (0)
2
= +00.00000 uA
Use
,
,
,
, ENTER or EXIT.
26. Set the Model 5700A current to 0 nA and press the
ENTER
key. The instrument will prompt for
the negative full-range current:
I(-FS)
2
= -20.00000 uA
Use
,
,
,
, ENTER or EXIT.
27. Set the Model 5700A current to -19.0000
μ
A.
28. Adjust the displayed current to match the Model 5700A current, and then press the
ENTER
key
to complete calibration of the present range.
29. Press the
EXIT
key to return to normal display.
30. Repeat steps 18 through 29 for the 200
μ
A to 20 mA ranges, using
as a guide. Be
sure to set the Model 6482 to the correct range using the RANGE
and
keys, and use the
corresponding Model 5700A currents for each range.