6482-901-01 Rev. A / August 2012
Return to
16-61
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 16: SCPI Command Reference
DIRection <name>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:DIRection <name>Control arm
bypass
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection <name>Control trigger bypass
Parameters
<name> = SOURce
Enable control source bypass
ACCeptor
Disable control source bypass
Query
:DIRection?
Query state of bypass
Description
This command is used to enable (SOURce) or disable (ACCeptor) control source
bypass. When enabled, operation will loop around the control source on the first
pass in the layer. After that, repeat passes in the layer are held up and will wait for
the programmed control source event.
INPut <event list>
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut <event list>Enable
event detectors
Parameters
<event list> =SOURceEnable Source Event Detector
DELayEnable Delay Event Detector
SENSeEnable Measure Event Detector
NONE
Disable all event detectors in Trigger Layer
NOTE
Each event in the list must be separated by a comma (i.e. trigger:input source, delay,
sense).
Query
:INPut?
Query enabled event detectors in Trigger Layer
Description
When TLINk is the selected Trigger Layer control source, and an event detector in
the Trigger Layer is enabled, operation will hold up at that detector until an input
trigger is received using the trigger link. When the event detector is disabled,
operation will not hold up. It continues and performs the appropriate action.
A Trigger Layer event detector is enabled by including the parameter name in the
event list for the INPut command. For example, to enable the Source Event
Detector and Measure Event Detector, send the following command:
:TRIGger:INPut SOURce, SENSe
The Delay Event Detector will be disabled since the DELay parameter is not
included in the above event list.
NOTE
To disable all the Trigger Layer event detectors, the NONE parameter must be sent alone
(i.e. trigger:input none). If it is listed with any of the other parameters, NONE will be
ignored.
ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:ILINe <NRf>Select input line; arm
layer
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe <NRf>Select input line; trigger layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
5
Line #5
6
Line #6
Query
:ILINe?
Query input trigger line